Sensitivity of the LAMP assay Figure 1 presents sensitivity of th

Sensitivity of the LAMP assay Figure 1 presents sensitivity of the toxR-based LAMP assay when testing 10-fold serial dilutions of V. parahaemolyticus ATCC 27969 DNA templates. A representative optic graph and

corresponding melting curve analysis for the real-time PCR platform and a representative turbidity graph for the real-time turbidimeter platform are shown in Figure 1A-1C, respectively. On the real-time PCR platform, for templates ranging in concentration from 4.7 × 105 to 4.7 × 101 CFU per reaction tube, the average Ct values selleck kinase inhibitor of six repeats ranged from 17.35 to 40.72 min, with melting temperatures consistently falling at around 83°C. No amplification was obtained for the 4.7 CFU and 4.7 × 10-1 CFU templates. Therefore, the detection limit of the toxR-based LAMP assay run in a real-time PCR selleck machine was approximately 47 CFU per reaction. In the real-time turbidimeter platform, the average Tt values fell between 34.43 and 49.07 min for templates ranging from 4.7 × 105 to 4.7 × 102 CFU per reaction tube. In two out of six repeats, amplification of the 4.7 × 101 CFU template occurred (Figure 1C). Therefore, the

lower limit of detection for turbidity-based real-time LAMP assay was 47-470 CFU per reaction. Figure 1 Sensitivity of the LAMP ABT-888 clinical trial assay when detecting Vibrio parahaemolyticus ATCC 27969 in pure culture. (A) A representative optic graph generated using the real-time PCR machine; (B) Corresponding melting curve analysis for samples in (A); (C) A representative turbidity graph generated using the real-time turbidimeter. Samples 1-7 corerspond to serial 10-fold dilutions of V. parahaemolyticus ATCC 27969 cells ranging from 4.7 × 105 to 4.7 × 10-1 CFU/reaction; sample 8 is water. Clomifene The two PCR assays used to test the same set of V. parahaemolyticus ATCC 27969 templates by using F3/B3 and toxR-PCR primers had the same level of sensitivity, approximately 4.7 × 103 CFU per reaction tube (data not shown), i.e., up to 100-fold less sensitive than the toxR-based LAMP assay. Quantitative capability of LAMP for detecting V. parahaemolyticus in pure culture Figure 2 shows the standard curves generated

when detecting V. parahaemolyticus ATCC 27969 in pure culture based on six independent repeats in both real-time PCR machine (Figure 2A) and a real-time turbidimeter (Figure 2B). On the real-time PCR platform, the correlation coefficient (r 2) was calculated to be 0.95. When run in the real-time turbidimeter platform, the toxR-based LAMP assay had an r 2 value of 0.94. Figure 2 Standard curves generated when detecting Vibrio parahaemolyticus ATCC 27969 in pure culture. (A) Based on six independent repeats in a real-time PCR machine; (B) Based on six independent repeats in a real-time turbidimeter. Detection of V. parahaemolyticus cells in spiked oysters The sensitivity of detecting V. parahaemolyticus ATCC 27969 cells in spiked oyster samples is shown in Table 3.

Comments are closed.